DeVilbiss iFill® Personal Oxygen Station Instruction Guide (Model 535D)
CAUTION–Federal (U.S.A.) law restricts this device to sale by, or on the order of a physician.
WARNING-Read and understand this instruction guide before operating this equipment.
MADE IN THE USA of U.S. and Imported Parts
DANGER–NO SMOKING
ES
Estación de Oxígeno Personal DeVilbiss iFill® Guía de Instrucciones
(Modelo 535D)
PRECAUCIÓN–La ley federal de EE. UU. limita la venta de este dispositivo a médicos o a personas que dispongan de la correspondiente orden
médica.
ADVERTENCIA-Antes de usar este equipo asegúrese de leer y comprender toda la información contenida en esta guía de instrucciones.
FABRICADO EN EE. UU. de partes nacionales e importadas.
PELIGRO–NO FUMAR
FR
Générateur d’oxygène Personnel DeVilbiss iFill® Guide de l’utilisateur
(Modèle 535D)
ATTENTION–En vertu de la loi fédérale américaine, cet appareil ne peut être vendu que par un médecin ou sur ordonnance de ce dernier.
AVERTISSEMENT-Veuillez lire et comprendre ce guide avant d’utiliser cet équipement.
FABRIQUÉ AUX ÉTATS-UNIS avec des pièces des États-Unis et des pièces importées.
DANGER–NE PAS FUMER

Table of Contents

Other manuals for DeVilbiss iFill 535D

Questions and Answers

  • J
    Jenna GutierrezSep 9, 2025
    How to fix no control panel lights on DeVilbiss Health Care Products?
    • A
      Adam FinleySep 10, 2025
      If there are no control panel lights or audible alarm when your DeVilbiss Health Care Products unit is first plugged in, check the following: 1. A tripped circuit breaker or faulty breaker in 535D or a blown fuse in 535I: Reset or replace the circuit breaker, or replace the fuse. 2. No AC power at the receptacle: Check the power source or try a different receptacle. 3. The AC wire harness is not connected to the main PC board: Connect the AC wire harness. 4. The light panel ribbon cable is loose or not connected to the main PC board: Reconnect the ribbon cable to the main PC board. 5. A defective DC power supply board: Replace the DC power supply board. 6. A defective light panel: Replace the light panel. 7. A defective main PC board: Replace the main PC board.
  • F
    frank63Sep 12, 2025
    Why is my DeVilbiss Oxygen Equipment showing excessive fill times?
    • W
      Willie JuarezSep 12, 2025
      If you're experiencing excessive fill times, ensure the iFill oxygen cylinder rotary selector is set to OFF. Also, check the fill times against the Typical Oxygen Cylinder Fill Times chart. Verify that the cylinder is connected properly. Try pressing the start/stop switch to return to Standby, then remove and reinstall the cylinder. Finally, press the start/stop switch again to begin filling the cylinder.
  • S
    Stephen DanielsSep 14, 2025
    How to fix hissing sound from iFill oxygen cylinder DeVilbiss?
    • K
      Kristin JacksonSep 14, 2025
      If you hear a hissing sound from the iFill oxygen cylinder, turn the rotary selector OFF. If the hissing sound continues, contact your DeVilbiss provider.
  • F
    Faith HodgesSep 16, 2025
    How to fix no indicator lights on DeVilbiss Oxygen Equipment?
    • L
      Lance WilliamsSep 16, 2025
      If there are no indicator lights on, check that the personal oxygen station power cord is properly plugged into the wall outlet and ensure the fuse in the Mains Power Entry Module is not open. If there are still no indicator lights, contact your DeVilbiss provider.

Summary

Important Safety Warnings

Danger - No Smoking

Critical safety warning indicating prohibition of smoking near the oxygen station.

Important Safeguards

General Dangers and Warnings

General safety precautions and potential hazards associated with the oxygen station.

Set-Up Procedures

Operation and Troubleshooting

Filling The iFill Oxygen Cylinder

Step-by-step instructions for filling an oxygen cylinder using the station.

DeVilbiss iFill Personal Oxygen Station Troubleshooting

Common problems and solutions for the DeVilbiss iFill Personal Oxygen Station.

Maintenance and Specifications

Cleaning and Maintenance

Procedures for cleaning the oxygen station filters and exterior cabinet.

Specifications

Technical specifications for the DeVilbiss iFill Personal Oxygen Station.

DeVilbiss iFill 535D Specifications

General IconGeneral
ModeliFill 535D
Oxygen Purity93% ± 3%
Power Consumption350 watts
Electrical Requirements120V, 60Hz
Storage Temperature-4°F to 140°F (-20°C to 60°C)
Product TypeOxygen Concentrator
Operating Temperature50°F to 95°F (10°C to 35°C)
Operating HumidityUp to 95% non-condensing
Alarm SystemPower failure, High temperature
Power SupplyAC