PPC2 AF OPERATION AND MAINTENANCE MANUAL
© 1998 DH Instruments, Inc. Page 100
OPERATION
To access the autozero, press 1AutoZ under 4Internal of the
SPECIAL menu. The display is:
Autozero
1View 2Edit 3Run
Selecting 1View allows the currently used values of PA(z)net and PA(z)tare (see Principle above) with the
date that the PA(z)net was determined to be viewed. Values for each range are stepped through by
pressing ENTER repeatedly. (See Section 5.2.3)
Selecting 2Edit allows the value of PA(z) net to be changed. (See Section 5.2.3)
Selecting 3Run leads to running the autozero sequence. (See
Sections 5.2.1 and 5.2.2) The display is:
Autozero ref sensor
1Hi 2Lo
Select 1Hi to run the autozero routine for the Hi (1 000 psia) reference transducer (see Section 5.2.1 Hi ref
by comp at atm).
Select 2Lo to run the autozero routine for the Lo (50 psia) reference transducer (see Section 5.2.2 Lo ref by vacuum).
Access for editing autozero values independently of the automated autozero routines is provided in
2Cal of 4Internal under the SPECIAL menu. Running the PA(z)tare autozero routine or editing and
activating PA(z)tare values causes the value of PA(z) net to go to zero. (See Section 5.4.4.5)
5.2.1 HI REF (BY COMP AT ATM)
PURPOSE
To automatically rezero (offset) the Hi reference pressure transducer (see Section 3.4.4.1).
PRINCIPLE
The autozero routine for the Hi reference pressure transducer uses the Lo reference pressure
transducer as a stable reference value against which to measure zero drift. This is valid since
the reference transducer zero drift is proportional to the transducer full scale and the full scale
of the Hi reference transducer is 20 times greater than the full scale of the Lo reference
transducer.
The Hi and Lo reference transducer readings are compared at atmospheric pressure and the
difference between the two is measured. The difference found [PA(z)raw], less the "natural"
difference [PA(z)tare], represents the zero drift since the last calibration of the Hi reference
transducer [PA(z)net]. The "natural" difference [PA(z)tare] is the difference found between
the two reference pressure transducers just after they have been calibrated.