PPC2 AF OPERATION AND MAINTENANCE MANUAL
© 1998 DH Instruments, Inc. Page 140
9.2 SPLT CABLE AND CONNECTIONS
In order to take advantage of the automatic purging feature of the PPC2 AF (see Section 3.2.8), the SPLT
valve must be connected to the PPC2 AF’s external driver #8. This can be accomplished by soldering two
24 gauge wires of desired length to the 12 pin connector provided with the accessories. Connect one wire
to pin L and the other to pin B. Attach the other end of these wires to the terminal block on the SPLT.
Polarity is not significant to the operation of the purge valve.