Rev. 1.2 (12.2021) MINI-CUBE | SERVICE MANUAL 24
Indicate the reference scale to be used for test results
(Westergren or Panchenkov).
Allows to reset and insert the temperature measured
inside the instrument.
Diagnostics on touchscreen, buzzer, movements,
transponders, EEPROM, SD card, readers, barcode
and Bluetooth. For the detailed description of the
tests, please see paragraph 3.1.3 “Diagnostics
Funtions”.
The instrument's serial number
To test the Bluetooth communication with an external APP
Rebuilt of the archives indexes. It may be required if some
settings were changed.
To set the screen save function and the brightness of the MINI-
CUBE written.
TIME STANDBY: time of inactivity, in seconds, after which the
instrument goes in Stand-by mode. Set = 0 to deactivate the Standy-by
management. Default setting = 600
BACKLIGHT: the level of the display’s backlight during the Stand-By
Mode. The value may vary between 0 and 255, default value is set to 120
TEST: to check the Stand-by mode.
You can decrease the number of available test on the instrument.
CAUTION: all removed test cannot be restored.
HELP IMPORT: Import from a USB key all the
.bmp
and
help_xxx.txt
files. NOTE: this function also imports the image of
initial presentation "hello.bmp".
HELP DELETE: Delete all
help_xxx.txt
and
.bmp
files.
NOTE: this function also erases the image of initial presentation
"hello.bmp".
ARCHIVES DELETE: Delete all archives.
LOG DELETE: Delete all log files.
SD BACKUP: Copy all internal MINI-CUBE files in a USB key. It is
advisable to do this operation in a completely empty key.
SD RESTORE: Copy all files in a USB key on the µSD inside the MINI-
CUBE
SD FORMAT: Formats the internal µSD of the MINI-CUBE.
WARNING: This operation involves the loss of all the internal files
of the MINI-CUBE.
EXPORT: Export the Configuration Parameters into a .TXT file on a
USB key
PRINT: Print the list of the Configuration Parameters using the
external Bluetooth printer.
DEFAULT: Set the Configuration Parameters to their default values.