What to do if Diesse cube 30 touch displays 'Used TEST DEVICE'?
- MmatthewwilsonSep 23, 2025
If the Diesse Laboratory Equipment displays 'Used TEST DEVICE', it means the test Device in use is empty. Obtain a new test device.
What to do if Diesse cube 30 touch displays 'Used TEST DEVICE'?
If the Diesse Laboratory Equipment displays 'Used TEST DEVICE', it means the test Device in use is empty. Obtain a new test device.
What to do if Diesse cube 30 touch Laboratory Equipment shows 'High (HI)' error?
The 'High (HI)' error indicates that the blood level inside the test tube is too high. If possible, remove blood until you reach the required levels.
What to do if Diesse Laboratory Equipment shows 'Low (Low)' error?
The 'Low (Low)' error means that the blood level in the test tube is too low. If possible, add blood until you reach the required levels.
What to do if Diesse cube 30 touch Laboratory Equipment displays 'FATAL ERROR! MOTOR FAIL!'?
This error indicates an engine problem with a component. In the message on the screen, press the 'OK' button, remove all samples from the plate, and restart the instrument. If the problem persists, contact Technical Assistance.
What does Err (2) (3) mean on Diesse cube 30 touch Laboratory Equipment?
Err (2) (3) indicates that the bottom of the test tube was not identified. Check that the test tube in use is compatible with the instrument and that it has been inserted correctly. Visually inspect the sample for presence of clumps or clots.
What does 'Error Removed tube Err (11)' mean on Diesse cube 30 touch Laboratory Equipment?
The 'Error Removed tube Err (11)' on the Diesse Laboratory Equipment appears at the end of the second reading if, during the cycle, the test tube is removed from its position. Repeat the test for that sample.
What does 'Err (2) (3)' mean on Diesse Laboratory Equipment?
The 'Err (2) (3)' error on the Diesse Laboratory Equipment indicates that the bottom of the test tube was not identified. Check that the test tube in use is compatible with the instrument and that it has been inserted correctly. Visually inspect the sample for presence of clumps or clots.
What causes 'Anomalous test tube Err (15)' on Diesse cube 30 touch?
The 'Anomalous test tube Err (15)' error on the Diesse Laboratory Equipment may be due to significant sedimentation with a low volume of blood. Check for the presence of clumps or clots. Try to repeat the test.
What causes 'Failed control method (13/14)' on Diesse Laboratory Equipment?
The 'Failed control method (13/14)' error indicates no confirmation of the control method. Check for any additional labels and position the tube as described in Chapter 4.
What to do when Diesse cube 30 touch says 'Exam not available Reloading required'?
This message indicates that there are zero tests remaining. Follow the instructions for reloading the instrument with a new test device.
Specifies the primary use and application of the CUBE 30 TOUCH instrument.
Provides an overview of the CUBE 30 TOUCH instrument, its design, and capabilities.
Highlights critical safety precautions and potential hazards associated with instrument use.
Specifies the required personal protective gear for safe operation and handling.
Describes the steps for safely cleaning and decontaminating the instrument.
Explains the procedure for updating the instrument's software to the latest version.
Provides guidelines for safely transporting and handling the instrument.
Outlines the complete steps for setting up and installing the instrument.
Details the procedure for replacing the instrument's fuses if they blow.
Describes the correct procedures for preparing blood samples for analysis.
Initiates the process of running an ESR analysis cycle on the instrument.
Details how to load test tubes into the instrument's carousel.
Explains the steps to begin an ESR measurement cycle after loading samples.
Describes how to add samples to the instrument during an ongoing test run.
Explains how to view and interpret the ESR analysis results.
Provides access to service menus and firmware information.
Details the process for registering quality control materials not supplied by the manufacturer.
Explains how to run quality control samples on the instrument.
Addresses issues and warnings related to operating temperatures outside the specified limits.