Rev. 2.3
CUBE 30 TOUCH | USER MANUAL
1
USER MANUAL
Rev. 2.3
Automated instrument for ESR determination
with modified Westergren method
Software version 2.xx.xx
FOR IN VITRO DIAGNOSTIC USE ONLY

Other manuals for Diesse cube 30 touch

Questions and Answers

  • D
    Dr. Jacob GarciaSep 23, 2025
    What to do if Diesse cube 30 touch displays 'Used TEST DEVICE'?
    • M
      matthewwilsonSep 23, 2025
      If the Diesse Laboratory Equipment displays 'Used TEST DEVICE', it means the test Device in use is empty. Obtain a new test device.
  • R
    rebecca97Aug 23, 2025
    What to do if Diesse cube 30 touch Laboratory Equipment shows 'High (HI)' error?
    • W
      williamortizAug 23, 2025
      The 'High (HI)' error indicates that the blood level inside the test tube is too high. If possible, remove blood until you reach the required levels.
  • W
    whitakerblakeAug 25, 2025
    What to do if Diesse Laboratory Equipment shows 'Low (Low)' error?
    • P
      proctortoniAug 26, 2025
      The 'Low (Low)' error means that the blood level in the test tube is too low. If possible, add blood until you reach the required levels.
  • A
    Andrew OwensAug 27, 2025
    What to do if Diesse cube 30 touch Laboratory Equipment displays 'FATAL ERROR! MOTOR FAIL!'?
    • C
      Cindy LeeAug 27, 2025
      This error indicates an engine problem with a component. In the message on the screen, press the 'OK' button, remove all samples from the plate, and restart the instrument. If the problem persists, contact Technical Assistance.
  • I
    ingramalexanderAug 27, 2025
    What does Err (2) (3) mean on Diesse cube 30 touch Laboratory Equipment?
    • J
      John HensonAug 27, 2025
      Err (2) (3) indicates that the bottom of the test tube was not identified. Check that the test tube in use is compatible with the instrument and that it has been inserted correctly. Visually inspect the sample for presence of clumps or clots.
  • J
    Jason EdwardsSep 17, 2025
    What does 'Error Removed tube Err (11)' mean on Diesse cube 30 touch Laboratory Equipment?
    • C
      chambersstacySep 18, 2025
      The 'Error Removed tube Err (11)' on the Diesse Laboratory Equipment appears at the end of the second reading if, during the cycle, the test tube is removed from its position. Repeat the test for that sample.
  • I
    ilopezSep 14, 2025
    What does 'Err (2) (3)' mean on Diesse Laboratory Equipment?
    • S
      Sandra HaleySep 14, 2025
      The 'Err (2) (3)' error on the Diesse Laboratory Equipment indicates that the bottom of the test tube was not identified. Check that the test tube in use is compatible with the instrument and that it has been inserted correctly. Visually inspect the sample for presence of clumps or clots.
  • D
    Douglas DyerSep 15, 2025
    What causes 'Anomalous test tube Err (15)' on Diesse cube 30 touch?
    • H
      htaylorSep 15, 2025
      The 'Anomalous test tube Err (15)' error on the Diesse Laboratory Equipment may be due to significant sedimentation with a low volume of blood. Check for the presence of clumps or clots. Try to repeat the test.
  • K
    kingkaylaSep 6, 2025
    What causes 'Failed control method (13/14)' on Diesse Laboratory Equipment?
    • S
      smithchaseSep 6, 2025
      The 'Failed control method (13/14)' error indicates no confirmation of the control method. Check for any additional labels and position the tube as described in Chapter 4.
  • J
    Jennifer RogersSep 3, 2025
    What to do when Diesse cube 30 touch says 'Exam not available Reloading required'?
    • A
      Aaron IrwinSep 3, 2025
      This message indicates that there are zero tests remaining. Follow the instructions for reloading the instrument with a new test device.

Summary

Introduction to CUBE 30 TOUCH

Intended Purpose of the Instrument

Specifies the primary use and application of the CUBE 30 TOUCH instrument.

Instrument Overview and Features

Provides an overview of the CUBE 30 TOUCH instrument, its design, and capabilities.

Safety Warnings and Precautions

Highlights critical safety precautions and potential hazards associated with instrument use.

Personal Protective Equipment (PPE) Requirements

Specifies the required personal protective gear for safe operation and handling.

Cleaning and Decontamination Procedures

Describes the steps for safely cleaning and decontaminating the instrument.

CUBE 30 TOUCH Technical Specifications

Software Update Procedure

Explains the procedure for updating the instrument's software to the latest version.

Instrument Installation Guide

Instrument Transport and Handling

Provides guidelines for safely transporting and handling the instrument.

Step-by-Step Installation Guide

Outlines the complete steps for setting up and installing the instrument.

Fuse Replacement Procedure

Details the procedure for replacing the instrument's fuses if they blow.

Operating the CUBE 30 TOUCH

Sample Preparation Guidelines

Describes the correct procedures for preparing blood samples for analysis.

Starting an ESR Analysis Cycle

Initiates the process of running an ESR analysis cycle on the instrument.

Loading Test Tubes into Carousel

Details how to load test tubes into the instrument's carousel.

Initiating the ESR Measurement Cycle

Explains the steps to begin an ESR measurement cycle after loading samples.

Adding Samples During Analysis (Random Access)

Describes how to add samples to the instrument during an ongoing test run.

Viewing and Interpreting Analysis Results

Explains how to view and interpret the ESR analysis results.

Instrument Settings and Configuration

Accessing Service Menu

Provides access to service menus and firmware information.

Quality Control Procedures

Registering Quality Control Materials

Details the process for registering quality control materials not supplied by the manufacturer.

Running Quality Control Samples

Explains how to run quality control samples on the instrument.

Troubleshooting Common Issues

Troubleshooting Temperature Issues

Addresses issues and warnings related to operating temperatures outside the specified limits.

Host Computer Connectivity

Diesse cube 30 touch Specifications

General IconGeneral
BrandDiesse
Modelcube 30 touch