CUBE 30 TOUCH | USER MANUAL
SARSTEDT S: MONOVETTE® 1,8mL
SARSTEDT S: MONOVETTE® 2,6mL
SARSTEDT S: MONOVETTE® 2,7mL
SARSTEDT S: MONOVETTE® 3,4mL
BD Microtainer® MAP (Microtube for Automated Process).
4.3 Test tube labelling
The CUBE 30 TOUCH’s optical sensors can detect blood sample levels in test
tubes with a maximum of one secondary label. Excessive thickness and
possible wrinkles on secondary labels, may increase the test tube’s external diameter,
increasing the risk of a tube becoming lodged in the instrument. If the operator notes
resistance when inserting a test tube, remove the added label before proceeding.
Secondary labels must adhere perfectly to the test tube’s surface to avoid possible label
fragments accumulating in the instrument, and thus obstructing analysis of the sample.
Secondary labels must be placed directly over the tube manufacturer label, leaving at
least 5 mm free space for the optical sensors to detect the blood level in the sample tube
(Figure 22).
When placing a label on a tube:
• Ensure the label is flattened smooth against the tube.
• Press the label down securely, including all the edges and corners to ensure that
no part of the label is loose.
• Leave a label-free space on the back.