CUBE 30 TOUCH | USER MANUAL
For QC samples the printing from the archive does not have any changes.
4.5.7 Interruption of a work session
In the CUBE 30 TOUCH it is possible to interrupt the work cycle at any time by pressing
the ‘STOP’ button. If the interruption is confirmed (Figure 48) the relative red status
string “Cycle xx canceled- Ready” is displayed and the analysis is aborted.
Figure 48 - Message to confirm the stop cycle
The processed samples are always stored in the archive but the string 'ERR!' will be
shown in the ESR column. In the sample detail, the error generated is "ERR (16)!", See
chapter 9.
If the cycle is interrupted before reading at time t = 0, the samples are saved in the
archive but not in the log file.