CUBE 30 TOUCH | USER MANUAL
11 PERFORMANCES
11.1 CUBE 30 TOUCH precision
Test Protocol: Sixty vials of ESR-Chex Plus control (30 vials of level 1 and 30 vials of level
2) were used in the study. The level 1 vials were inserted in each well of the instrument
and tubes were analyzed 3 times over the course of the 7-day open-vial stability of the
control. The study was repeated with level 2 vials. The mean, standard deviation and
coefficient of variation (CV%) were calculated for intra-assay precision (repeatability and
within lab).
ESR Value (mm/h)
Level 1, Lot 403
ESR Value (mm/h)
Level 2, Lot 403