CUBE 30 TOUCH | USER MANUAL
Caution! “N” tests remain
The number of remaining
tests is less than 50.
Be sure to reload the device
as soon as possible with a
new test device
The test Device in use is
empty.
Obtain a new test device.
FATAL ERROR!
MOTOR FAIL!
Xxxxxx
yyyy
An engine error on
component xxxxx has
occurred (ex. Reader, Plate
etc.). The type of error is
indicated by yyyy (e.g.
Home, step, etc.)
n the message that appears
on the screen, press the 'OK'
button. Remove all samples
from the plate and restart
the instrument.
If the problem persists,
contact Technical
Assistance
9.1 Temperature out of range
If the working temperature of the instrument does not respect the 15 °C - 35 °C limits
established in the technical characteristics (see paragraph 2.4), the condition is signaled
to the operator as soon as the 'Start' button is pressed to start a cycle analysis (Figure 71).
Figure 71 - Temperature out of range
To proceed with the analysis, the operator must confirm, otherwise the instrument does
not start processing the samples.
ATTENZION: The results of samples processed with temperatures outside the
range of 15-35 ° C (operating temperature range) may not comply with the declared
performance.