1.8.1 Cleaning and decontamination procedure
External cleaning is recommended for safety reasons.
The decontamination procedure must be carried out by the user in the event of
biological material leakage, displacement of the instrument or when deemed necessary.
In the event of biological material leakage during the operating cycle, clean the
external surfaces of the instrument with personal protective equipment.
Decontamination procedure:
1. With the instrument turned off, clean the instrument with a disinfectant liquid
used in the laboratory and allow it to dry. For the touch screen use a dry microfiber
cloth.
2. Leave the instrument turned off for at least 1 hour before starting a new operating
cycle or performing any other operation on the instrument.
For assistance with the inside of the instrument not easily accessible to the operator,
contact Technical Services.