7 QUALITY CONTROL
Quality Control materials supplied by DIESSE Diagnostica Senese S.p.A are reported
among the instrument accessories. They are automatically and conveniently identified
by the instrument. For this reason, registration of the control material is not necessary.
QC vials may be run with or without the plastic adaptor.
In case of use of Quality Control materials not supplied by DIESSE Diagnostica Senese
S.p.A, refer to section 7.1 for the registration procedure.
7.1 QC registration procedure
To utilize quality control material not supplied by DIESSE Diagnostica Senese S.p.A:
1. Apply dedicated barcode labels
2. Select the item QC from the Settings menu (Figure 56 then Figure 60)
3. Insert the Normal control in position 1 and press the DETECT icon. The instrument
detects the code and displays it in the barcode window
4. Manually insert the relative data for name, lot, expiry, and acceptability limits
5. Remove the Normal QC
6. Insert the Abnormal QC in position 1
7. Repeat the same procedure as at point 3
Following this procedure, the instrument will be able to identify the controls
automatically from the barcode.