Rev. 2.3
23
Figure 11 - Instrument extraction
5. Remove the protective bag which wraps the instrument.
6. Verify that the packing list and supplied materials match.
3.5 Environment
For normal safety requirements and given the type of testing it performs, the instrument
must be positioned in a dust-free environment away from heat sources, and free from
any exposure to liquids. Position the instrument on a perfectly level bench, not subject
to shaking or vibrations. Observe a 15 cm or 6” perimeter around the instrument as a
safety precaution, as indicated in the Figure 12. In addition, it is advisable to position the
instrument a 1 meter distance from devices that generate electromagnetic waves (e.g.
laboratory refrigerators, centrifuges).
Figure 12 - Instrument placement
The safety of the instrument and of the operator is not guaranteed if one or more of the
following conditions are violated.
The mains must be compatible with the voltage and current specifications
indicated on the metal plate affixed on the rear of the instrument.
Verify the compatibility of external accessories such as a BC reader or USB devices
prior to connecting.