THRESHOLD VALUE: Set a threshold value; when this value is exceeded the
instrument will display the result in red to alert the operator that the result is outside
of the normal range. The default value is 20 mm/hr.
The threshold value that can be set from this screen is automatically attributed to all
types of tubes; the modification on the single type can be made in the 'Service' , refer
to the Service Manual.
QC DIESSE DECODE: Deactivating (OFF) this function, the instrument will NOT
recognize (in automatic mode) a DIESSE control by reading only the label applied on
the tube. In this case, the QC must first be registered as described in the registration
procedure in Chapter 5 to be recognized.
DEFAULT TUBE TYPE: allows the user to choose a default tube type. With this
selection, any type of tube inserted will be managed as the type of tube set. The types
of test tubes that can be chosen are those compatible with the instrument already
presented in paragraph 1.1.
The default setting is 'NONE': the instrument automatically detects the type of tube
inserted.
TUBE TYPE SET PAUSE: enables the message window to confirm or modify the type
of test tube detected by the instrument. If the setting is disabled, the type of tube is
automatically confirmed, and the cycle is started automatically.
QC: Enter the QC menu to configure quality control settings (See chapter 5).
LOG EXPORT: Export the instrument’s Log file to a USB device, in Log.txt.
Enter the range of dates of interest, press the 'OK' button, insert the USB device into
the port (Figure 7), wait for the operation to complete. The exported files follow the
following nomenclature: yyyymmdd.LOG where “yyyymmdd” represents the date of
the working day. If, in the selected date range, there is no work cycle, a file called
xxxx_RDP.txt is saved (where xxxx are the last 4 digits of the serial number)
containing the parameters of the instrument.
ARCHIVES EXPORT: Export the data (SAMP.CSV) or QC(QC.CSV) archive to a USB
device, in *.xls format. For each sample the following information is present: the
barcode, date and time of the examination, the position inside the sample plate, the
ESR value (ESR), FLAG, temperature and cycle number. The QC file reports: the name,
the lot, the expiry date, the minimum and maximum ranges, the type, date and time
of the processing, position in the plate and the ESR value.
ARCHIVES BACK UP: Back up or restore the archives. The files are stored in the
complex format .arc: dbsamp.arc, dbqc.arc.