AP Dynamometer Service Manual
Distributed by:
ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772
eFax: +1 (866) 234-0451
Web: http
s
://www.abqindustrial.net E-mail:
info@abqindustrial.net

Table of Contents

Other manuals for Dillon AP

Questions and Answers

  • J
    James LesterJul 28, 2025
    How to fix nonlinear readings on Dillon Measuring Instruments?
    • T
      Tina ReynoldsJul 28, 2025
      Nonlinear readings on your Dillon Measuring Instruments can be caused by a bent bar or worn gauge movement. Try replacing the pressure bar or the gauge movement to correct this issue.
  • K
    Kerry ClarkeAug 3, 2025
    What to do if Dillon AP Measuring Instruments is high on low end, low on high end?
    • O
      odixonAug 3, 2025
      If your Dillon Measuring Instruments reads high on the low end and low on the high end, it could be due to a bad pressure arm. Replacing the gauge movement is the suggested solution.
  • O
    Olivia ReynoldsAug 9, 2025
    Why does the pointer jump in my Dillon Measuring Instruments?
    • J
      Jonathan MillerAug 9, 2025
      If the pointer jumps, it might be because the bearings or pointer are rubbing on the dial face. To resolve this, try cleaning the bearings, applying one drop of LPS on the bearing, or replacing the gauge movement.
  • B
    Briana FryeAug 15, 2025
    What to do if max hand will not rotate smoothly on Dillon AP Measuring Instruments?
    • D
      Donna Thomas DVMAug 15, 2025
      If the max hand doesn't rotate smoothly, the spring might be too tight or rusty/corroded. The suggested solution is to replace the max hand assembly.
  • W
    Wayne JonesAug 21, 2025
    Why Dillon Measuring Instruments Won’t repeat?
    • L
      Laurie PriceAug 21, 2025
      If your Dillon Measuring Instruments won't repeat, it could be due to bad bar bearings that are worn, rusty, or dirty. Replacing the bearings in the pressure bar might solve this. Another possible cause is a worn gauge movement, in which case you should replace the gauge movement.

Summary

Disassembly and Reassembly

Separation of Pressure Bar and Gauge Movement on 5 Diameter Units up to 20,000 lb;10000 kg

Steps for removing the pressure bar from 5" diameter AP dynamometer units.

Reassembly of Pressure Bar and Gauge Movement on 5 Diameter Units up to 20,000 lb;10000 kg

Steps for reassembling the pressure bar on 5" diameter AP dynamometer units.

Separation of Pressure Bar and Gauge Movement on 10 Diameter Units up to 20,000 lb;10000 kg

Steps for removing the pressure bar from 10" diameter units up to 20,000 lb.

Reassembly of Pressure Bar and Gauge Movement on 10 Diameter Units up to 20,000 lb;10000 kg

Steps for reassembling the pressure bar on 10" diameter units up to 20,000 lb.

Periodic Maintenance Requirements

Calibration Procedures

Troubleshooting and Common Repairs

Dillon AP Specifications

General IconGeneral
CategoryMeasuring Instruments
TypeAnalog Pressure Gauge
Accuracy±1% of Full Scale
DisplayAnalog dial
PowerNone (mechanical)
CapacityVaries by model
Units of Measurepsi, bar, kPa
Dial Size2.5 in or 4 in (depending on model)
Connection1/4" NPT or 1/2" NPT (depending on model)
MaterialBrass or Stainless Steel (depending on model)
Liquid FilledOptional