EDXtreme User Instructions 23
Auto ID Increment: Enable/Disable
This will increment the ID by 1 every time the
"PRINT" key is pressed and store it with each
force and peak.
Comm Port
:
This allows you to choose where to export the
"ON PRINT" logged data.
a. EDX = Cell
b. COMM = Com1
c. RADIO = Future development
d. Disabled = to turn off
Timed: This is used when the operator wants to record a
live load/pull and store it internally. Can store up to 255
force readings.
Log Rate
:
The speed at which the logging of force will be
recorded. Based on seconds (Example 5 sec =
Unit will record the force once every 5 seconds)
Log Count
:
This is how many logs the operator wants to
record. Range is 0 to 255 recorded readings.
Auto Send
:
Can export the readings live if enabled.
Comm Port
:
This allows you to choose where to export the
"ON PRINT" logged data.
a. EDX = Cell
b. COMM = Com1
c. RADIO = Future development
d. Disabled = to turn off
Overload: This is used when an operator or supervisor
wants to know if their equipment is being overloaded.
This will be a percentage based on capacity and will be
below 100% of the EDX. (This not the same as the 120%
overload that the EDX records based on capacity.)
Maximum Records = 255
Threshold %
: (based on capacity)
This the percentage above where the unit will
record the peak load.
Re-arm %
: (based on capacity)
This is the load percentage the force must drop
below before the "OVERLOAD" will reset (re-arm)
and store another load.