Electromagnetic immunity (EMI)
49
Table 2
Explanation – electromagnetic immunity
The device is intended for use in the electromag-
netic environment specified below. The customer
or the user of the device should ensure that it is
used in such an environment.
Immunity
tests
IEC 60601
test level
Com-
pliance
level
Electromagnet-
ic environment
– instructions
Electro-
static
discharge
IEC
61000-4-2
± 8 kV
contact
± 2 kV,
± 4 kV,
± 8 kV,
± 15 kV air
± 8 kV
contact
± 2 kV,
± 4 kV, ±
8 kV, ±
15 kV air
Floors should
be wood, con-
crete or ceramic
tile. If floors
are covered
with synthetic
material, the
relative humid-
ity should be at
least 30 %.
Fast
transient
electrical
distur-
bances/
bursts
IEC
61000-4-4
± 2 kV for
power
supply
cables
± 1 kV for
input/out-
put cables
Not ap-
plicable
Outlet power
quality should
be that of a typ-
ical commercial
or hospital
environment.