Electromagnetic immunity (EMI)
52
Immunity
tests
IEC 60601
test level
Com-
pliance
level
Electromagnetic
environment
– instructions
Conduct-
ed RF
IEC
61000-4-6
3 V
0.15 MHz
to 80MHz
6 V in ISM
and ama-
teur radio
bands be-
tween 0.15
MHz and
80 MHz
Not ap-
plicable
Portable and
mobile RF com-
munications
equipment
should be used
no closer to
any part of the
device than the
recommended
separation
distance cal-
culated from
the equation
applicable to the
frequency of the
transmitter.
Recommend-
ed separation
distance
150 KHz to 80
MHz
80 MHz to 800
MHz
80 MHz to 2.7
GHz
where P is the
maximum
output power
rating of the
transmitter
in watts (W)
according to
the transmitter
manufacturer
and d is the rec
-
ommended sep-
aration distance
in meters (m).
Radiated
RF
IEC
61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m