Ensure that label indicated the correct voltage before connecting the instrument
to power supply
Do not operate the unit with a damaged power cord
Set-up and install the magnetic stirrer on a stable and temperature resistant
surface
Ensure good working environment free of explosive, hazardous and
inflammable substances or water.
Ensure that the main power supply cable does not touch the hotplate.
Keep the magnetic Stirrer away from the effects of high magnetic field.
Do not cover the instrument and do not splash water on the components
When working wear personal safety guards to avoid the risk of splashing and
evaporation of liquids, release of toxic or combustible gases.
Heating temperature must be set to at least 50 °C lower than the fire point of
the chemicals used.
Pathogenic materials processing must be done in closed vessels only.
Ensure that the tip of the external temperature sensor is at least 5-10 mm away
from the bottom of the vessel.
Always disconnect the main power supply after use.