Warning!
Forbid to transfer the vessel when the instrument working.
4. Function
4.1. Heating(hotplate stirrer/hotplate)
The instrument is controlled by digital temperature control technology, which has
two separate safety circuits. The hotplate is kept at a constant temperature by a
digital control circuit. The hotplate temperature can also be monitored from a
separate, adjustable safety circuit. The two temperature sensors PT1000 internal
builds temperature control into the hotplate and PT1000 external can monitor the
temperature of sample.
Set the target temperature and start heating:
Set the temperature by rotating the temperature control knob slowly to desired
setting.
The set value is confirmed and heating is switched ON by pushing the temp.
Control knob.
The LED displays the temperature value on the left-hand side during heating.
The heating function is switched OFF by pushing the same temperature control
knob again.