What to do if Ernn appears on Doran DS6200 display?
- PPaula RoachAug 7, 2025
If 'Ernn' appears on the display of your Doran Scales, it indicates excessive motion on the scale platform. Ensure the platform is stable and free from vibrations.
What to do if Ernn appears on Doran DS6200 display?
If 'Ernn' appears on the display of your Doran Scales, it indicates excessive motion on the scale platform. Ensure the platform is stable and free from vibrations.
What to do if ErSp appears on Doran DS6200 Scales display?
If 'ErSp' is displayed on your Doran Scales, place the correct weight on the platform. Then, press the BMI button to perform the span calibration again. If the error continues, consult the raw counts table.
How to fix gruL or groL error on Doran Scales?
If 'gruL' or 'groL' appears on your Doran Scales display, check the connections between the indicator and the base. Then, press the BMI button to perform the zero calibration again. If the error persists, consult the raw counts table.
Carefully remove the scale from the shipping carton and notify the shipper of any damage.
Observe precautions to ensure years of trouble-free service and avoid damage.
Tips for caring for the scale to ensure longevity, including approved cleaning methods and liquids.
Periodically examine the scale for damage, check calibration, and calibrate if necessary.
Disables the scale during shipping by pressing and holding the ZERO button for 5 seconds.
Charge the scale using the included USB A to USB C cable and a 5V/1A adapter.
The lowbatt annunciator flashes when the battery is low; the scale operates while charging.
Overview of front panel pushbuttons and display annunciators for scale operation.
Covers powering on/off, zeroing, unit selection, reweigh, and BMI input.
Instructions for resetting the scale and viewing software revision.
Step-by-step guides for weighing patients, mother/baby, and BMI calculations.
Procedure to enter setup mode and access calibration options.
Details on Two Point and Multiple Point calibration methods for accuracy.
Troubleshoots calibration errors like zero out of range, span out of range, and motion detection.
How to enter, exit, and navigate through the parameter setup menu.
Options to enter Two Point or Multi Point calibration modes from the setup menu.
Option to reset all parameters to factory defaults, preserving calibration data.
Controls the time the scale stays on after non-use; extending timeout reduces battery life.
Determines motion allowed for stable weight; slower setting for accuracy, faster for speed.
Controls information transmission via RS232 or USB port, including EMR data.
Specifies the format protocol for transmitted and received data from the serial port.
Enables or disables pound (lb), kilogram (kg), inches, and centimeters units.
Enables or disables the indicator beep when buttons are pressed.
Enables or disables beep when a stable or average hold weight is achieved.
Displays current battery voltage (DC) with indicator lights for charge status.
Raw counts from A/D converter; view for troubleshooting calibration issues.
Displays the current software revision of the scale.
Disables scale when in transport; exit by pressing and holding ZERO for 5 seconds.
Indicator has a USB C port for EMR Communication Module connection.
Defines commands and responses for Doran F1 data communication protocol.
Defines commands and responses for IQ Vitals data communication protocol.
Defines commands and responses for Welch Allyn Spot data communication protocol.
Provides solutions for common problems like incorrect readings, zero issues, and scale not turning on.
The Doran DS6200 is a remote indicator scale designed for professional use, particularly in healthcare environments. It features an internal rechargeable LiFePO4 battery, offering over 5500 weighments between charges, and can be charged using a USB A to USB C charging cable with any 5V, 1A or higher USB wall mount adapter. The scale is built to ISO9001 standards, ensuring high quality and reliability.
The DS6200 functions as a physician scale, providing accurate weight measurements for patients. It includes features for Mother Baby operation (allowing an infant's weight to be measured while held by a parent after the parent's weight is tared out) and BMI calculation (requiring manual height entry). The scale automatically turns on when a patient steps on it and captures a stable weight. It can also be manually turned on by pressing the ZERO button. An adjustable Automatic Shutoff Timer helps conserve battery life by turning off the indicator when not in use. The scale supports both pounds (lb) and kilograms (kg) weighing units, with a Units Lock feature available in parameter setup to disable unwanted units.