Section 5. Parameter Setup
The indicator has many adjustable parameters which control the scale’s operation. The
parameter settings shown in bold are the default settings.
Entering and Exiting Setup Mode
To enter setup mode from the front panel
1. Press ZERO to turn on the scale.
2. Press and hold ZERO and UNITS simultaneously for 5 seconds until the
display reads Prset.
3. Quickly press REWEIGH 3 times and the display will read CAL n.
To exit the setup mode and save changes, press REWEIGH or BMI to scroll to
the end of parameter menu. The display will read SAVED to confirm changes are
saved.
To exit the setup mode and abort all changes, wait 120 seconds after any key
press and the scale will exit setup mode and not save any changes.
Setup Mode Navigation
1. Press BMI to scroll forward to the next parameter.
2. Press ▲ or ▼ buttons to change the current parameter setting.
3. Press Reweigh to save any changes and exit setup mode.
Calibration Mode Entry
Enter Two Point Calibration
Do not enter calibration
Enter Calibration
(see Scale Calibration section)
Enter Multi Point Calibration
Do not enter calibration
Enter Calibration
(see Scale Calibration section)
Reset All Parameters to Default Settings
Default parameters
No. Parameters are not defaulted.
Yes. Set all parameters to the factory
default values when BMI is pressed.
Calibration data is preserved.