Multiple Point Calibration Procedure
1. When the setup mode is entered the display reads CAL n.
2. Press BMI to navigate to CAL nn
3. Press the ▲arrow button to change the display to CAL ny.
4. Press BMI.
5. The display reads CAL 0.
6. Remove all weight from the scale platform.
NOTE: During calibration be sure vibration and air currents are not present.
Always use calibrated test weights.
7. Press BMI to perform a zero point calibration. The display will count down to
zero. The counter will reset if too much motion (Er nn) is detected.
8. The scale will display 100 lb (45 kg)
9. Place 100 lb (45 kg) onto the scale.
10. Press BMI. The display will count down to zero.
11. The scale will display 200 lb (90 kg)
12. Place 200 lb (90 kg) onto the scale.
13. Press BMI. The display will count down to zero.
14. The scale will display 300 lb (135 kg)
15. Place 300 lb (135 kg) onto the scale.
16. Press BMI. The display will count down to zero.
17. The scale will display 400 lb (180 kg)
18. Place 400 lb (180 kg) onto the scale.
19. Press BMI. The display will count down to zero.
20. The display reads SAVED and exits to the weigh mode. Verify calibration
accuracy with your test weights.