Automatic Shutoff Timer
Automatic shutoff timer. The timer
controls the period of time the scale
stays on after non-use. Extending
timeout will reduce battery life.
10 seconds (Default)
30 Seconds
1 minute
2 minutes
5 minutes
10 minutes
15 minutes
Automatic shutoff timer disabled.
Recommended to plug scale in
continuously.
Average Aperture
Determines the amount of motion allowed
to hold a stable weight. A slower setting
will result in more accurate weights but will
require a longer time to achieve a stable
weight. A faster setting will achieve a
stable weight faster but may be less
accurate.
Fastest Setting
Default Setting
Slowest Setting
Data Output
Controls when information is
transmitted from the scale though the
RS232 serial port and or USB port.
EMR is transmitted when a request
command is received through the or USB
port.
Transmits an EMR string after the
patient’s weight is displayed. (default)
Transmits an EMR string after the Height
has been entered
Continuously transmits an EMR string 3.75
times a second.