EasyManuals Logo

Dostmann Electronic P700 Series User Manual

Dostmann Electronic P700 Series
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
16
By displaying the following error codes the instrument
support the operation of the instrument.
Error Meaning
oPEn no probe
or wrong probe is connected
|
-
oLo
“too low“ – underflow of the
measuring range
|
-
oHi
„too high“ – exceeding of the
measuring range
|
-
Er1
temperature of the cold
junction above the measuring
range
|
-
Er2
temperature of the cold
junction above the measuring
range
E15 low battery – (battery is com-
pletely exhausted)
E19 low battery – (battery is com-
pletely exhausted)
E1d low battery – (battery is com-
pletely exhausted)
E16 time out / auto-off-function
E1o time out / auto-off-function
E1e time out / auto-off-function
E23 EE-prom destroyed or
EE-prom index is wrong
E25 EE-prom destroyed or
EE-prom index is wrong
E31 temperature of cold junction
out of range
3. Error Codes