32 Dräger Alcotest 9510 Washington Technical Manual V3.1 2014
Technical Data
17 Technical Data
Resolution: 0.001%BrAC (0.001 g/210 liters of breath)
Measurement Range: 0.000% to 0.630% BrAC
Operational Conditions
Temperature Range: 0°C to 40°C; 32°F to 104°F
Relative Humidity: 10% to 99% relative (non condensing)
Atmospheric pressure: 600 hPa to 1050 hPa
Elevation: Maximum height of 3000 meters.
Storage Temperature Range: -20°C to 70°C; -4°F to 158°F
Warm-up Time: Less than 15 minutes at 20°C
Printer: High speed, high resolution thermal printer. Standard paper 2 1/4” wide (58 mm) and 85’ long
Date/Time Display: Month/Day/Year and Hour:Minute (24 hour clock)
Calibration Interval: Recommended interval for verifying accuracy of instrument a
nd, if required, r
ecalibration: 12 months
Standard compliance: NHTSA, OIML R 126 : 1998
Electrical Characteristics
Operating voltages
AC power: 110 - 240 VAC, 50/60 Hz
DC power: 10.5 - 15.5 VDC
Fuse: T 2A H, 250 V; IEC 60127-2/V (two fuses)
Power Consumption
During warm-up: approx. 70 Watts
During measurement: approx. 30 Watts
Stand-by: < 15 Watts
Dimensions: 12.9” x 9.8” x 2.2” (front) 7.3” (back) (W x H x D)
Weight: Approximately 15.3 lbs
CE-Markings: Directive 2004/108/EC for ele
ctr
omagnetic compatibility
Low Voltage Directive (LVD) 2006/95/EC