INFINITY COCKPIT QUICK REFERENCE GUIDE |30
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
o The computation constant is already entered except Other.
a) If Other is selected, manually enter computation con-
stant by selecting Comp. Constant [ ]
b) Using adjacent on-screen keypad or keyboard, touch
to enter numbers.
c) Click rotary knob to confirm.
6. Select Size
o Dial rotary knob to select desired setting.
o Click rotary knob to confirm.
TO PERFORM A CARDIAC OUTPUT MEASUREMENT
1. Select C.O. parameter box.
o If necessary, select Mode Auto or Manual.
2. Select Start C.O.
OR
3. Select the Procedures... button from the main menu bar.
4. Select the C.O. tab if necessary.
5. Select Start C.O.
o In Auto mode, the message Inject when ready appears in
the Procedures... > C.O.
a) Blood temperature must be stable (from previous in-
jectate solution), otherwise will give error message.
o In Manual mode, process is as above, but no Inject when
ready messaging.
a) Inject as desired.
6. To eliminate inconsistent values from being included, touch
any of the results to eliminate them from the average.
7. Before leaving the procedures page, select the Save C.O. av-
erage button to store it for calculations.