INFINITY COCKPIT QUICK REFERENCE GUIDE |16
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
Alarm Pause & Alarm Setup
TO PAUSE AUDIO ALARMS FOR 2 MINUTES
1. Press the yellow key on the Cockpit.
2. Pressing the key again turns alarms back on
Note: Quiet Mode is Activated – If a new alarm condition with a priority
higher than the currently paused alarm occurs, a truncated alarm tone
sounds. In addition, the alarm is represented by visual alarm signals corre-
sponding to the alarm priority. If the new alarm is of lower priority than the
paused alarm, the new alarm condition is only represented by a visual alarm
signal. No acoustic alarm tones sound.
If Quiet Mode is Deactivated - Any new alarm condition breaks through the
Audio Pause period with full acoustic and visual alarm annunciation.
TO PRE-SILENCE ALARMS FOR 2 MINUTES
This function allows you to pre-silence (audio pause in advance)
potential alarm conditions before they occur.
1. Press the yellow key on the Cockpit
o Alarm conditions are reported visually by a corres-
ponding alarm message and a blinking parameter box.
o A single alarm tone sequence is generated for the first
occurrence of an alarm condition of low, medium or high
priority.
o The Audio paused % banner appears in the far right
field of the header bar along with a countdown timer
and the following symbol:
o If multiple alarm conditions arise during an active pre-
silence period, the Cockpit triggers a single alarm tone
sequence for the highest grade.
Note: Pre-Silence functions only if Quiet Mode is activated.