2015 Draeger
Medic
al
,
Inc
.
INFINITY COCKPIT QUICK REFERENCE GUIDE |17
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
ARCHIVE FUNCTION
In response to an alarm limit violation for each Parameter or Ar-
rhythmia the following occurs depending on the Archive setting:
o Off – no event is stored and no recording is generated.
o Store – stores the event for later review.
o Record – generates a timed recording
o Str/Rec – generates a timed recording and stores the event.
Note: All Arrhythmia Alarms if set to Str or Str/Rec will hold the event in the
alarm history for later review even when the alarm is turned Off.
CONFIGURING INDIVIDUAL PARAMETER ALARM SETTINGS
1. Select desired parameter box to access the parameter setup
page.
2. Select button above the current patient value to adjust the
High limit. Select the button below the current patient value to
adjust the Low limit.
3. When button is highlighted yellow use Rotary knob to adjust
the parameter limit higher or lower.
4. Press Rotary knob to confirm.
5. Select button below low limit to select how an alarm violation
for this parameter will be Archived.
6. To turn alarm Off or On for the parameter
o Select the Alarm on/off button, a crossed-out triangle
appears in the parameter box when alarm monitoring is
deactivated.
WARNING: The banner HR alarms off appears in the right most
field in the header bar whenever you deactivate heart rate alarms.
Setting alarm limits to extreme values may prevent certain alarm
conditions from being detected and from being annunciated with
acoustic and visual alarm signals.