NM2B ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
5.5.17 The nitrous oxide flow should stop when the flow of oxygen is
between 250 and 400 ml per minute.
5.5.18 Close the N
2
O flow control valve and turn the System Power switch
to STANDBY.
5.5.19 Reconnect the pipeline hoses.
5.5.20 Replace the flowmeter housing rear cover.
5.5.21 Perform the PMS Procedure given in Section 6.
Rev. D
5-17