PMC PROCEDURE (continued) NM2B
( ) 6.8.3 Lubrication, Spiromed Sensor
6.8.3.1 Locate the four lateral holes at the sides of the Spiromed sensor
marked by printed arrows.
CAUTION: Use only Sensor Lubrication Kit P/N 2218180 for the
following procedure.
6.8.3.2 Remove the protective cover from the pipette and open the
lubricant bottle.
6.8.3.3 Dip the tip of the pipette into the lubricant and draw one drop of
lubricant into the pipette by pulling the pin upwards.
6.8.3.4 Insert the pipette into one of the four holes as far as it will go.
Push the pin forward to its stop to inject lubricant into the hole.
6.8.3.5 Repeat the procedure on the three remaining holes.
6.8.3.6 Wipe any lubricant residue from the exterior of the sensor.
6.8.3.7 Reattach the sensor to the absorber top dome.
6.8.3.8 Reattach the expiratory valve to the sensor.
Rev. Y
6-16