TROUBLESHOOTING GUIDE (continued) NM2B
GUIDE 5: Vitalink Failure
N
Y
Y
N
N
Y
N
Y
N
Y
START
UNIT IS
FUNCTIONAL
DOES UNIT
PASS PMS?
PERFORM A
COMPLETE PMS
ON UNIT
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
DOES
HOST FAIL TO
COMMUNICATE WITH
PERIPHERAL
DEVICE?
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
DOES
HOST FAIL TO
COMMUNICATE WITH
PERIPHERAL
DEVICE?
CONNECT
PERIPHERAL CABLE
TO SERIAL
CHECK PERIPHERAL
& HOST SERIAL
PORT
CONFIGURATION
SETTINGS
IS
PERIPHERAL
CABLE CONNECTED
TO SERIAL INTERFACE
PORT?
DOES
HOST FAIL TO
COMMUNICATE WITH
PERIPHERAL
DEVICE?