TROUBLESHOOTING GUIDE (continued) NM2B
GUIDE 3: Loss of Respiratory Volume Monitor
N
Y
N
Y
N
Y
N
Y
Y
N
N
Y
START
UNIT IS
FUNCTIONAL
DOES UNIT
PASS PMS?
PERFORM A
COMPLETE PMS
ON UNIT
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
IS
VOLUME MONITOR
FUNCTIONAL?
REINSTALL ORIGINAL
VOLUME SENSOR
------------------------
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
IS
VOLUME MONITOR
FUNCTIONAL?
CONNECT CABLE TO
J9 ON PROCESSOR
BOARD (ALSO
CHECK CONTINUITY OF
INTERNAL CABLE)
CONNECT VOLUME
SENSOR CABLE TO
INTERFACE PANEL
REPLACE
VOLUME SENSOR
ASOUTLINED IN
PROCEDURE 4.29
IS CABLE
CONNECTED TO
J9 ON PROCESSOR
BOARD?
IS
"VOL SENS DISC"
DISPLAYED ON
CENTRALERT ALARM
DISPLAY?
IS
VOLUME MONITOR
FUNCTIONAL?
Rev. A
3-6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS