PMC PROCEDURE (continued) NM2B
6.5 SERVICE DATA
6.5.1 Access the Configure screen, then press the following keys on the
monitor controls key panel: 21% key and APNEA ALARM DISABLE
key (simultaneously), and then the CONFIG key (while still holding the
previous two keys). The alarm CRT should then display the Service
Menu Screen.
6.5.2 Press the DIAGNOSTICS key. The alarm CRT should display the
Diagnostics Menu Screen.
( ) 6.5.5 Record any Error Log codes.
( ) 6.5.3 Record the Last Service Date.
( ) 6.5.4 Press the RESET DATE key.
6.5.6 Key Panel
6.5.6.1 Press the KEY TEST key.
6.5.6.2 The alarm CRT should display the outline of all keys on the
display panel.
6.5.6.3 Press each key on the display panel, one at a time.
( ) 6.5.6.4 As each key on the display panel is pressed, do all the
corresponding keys on the alarm CRT illuminate? ___(Y)
NOTE: The TREND key should be pressed last, because it also exits the
Key Panel Test Screen.
Rev. Y
6-6