EasyManua.ls Logo

Dräger Narkomed 2B - A Bellows: Pediatric External

Dräger Narkomed 2B
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PMC PROCEDURE (continued) NM2B
6.25A BELLOWS: PEDIATRIC EXTERNAL - if applicable
6.25A.1 Set the ventilator frequency to 10 BPM.
6.25A.2 Set the I:E Ration to 1:2.
6.25A.3 Adjust the tidal volume to 300 ml/min.
6.25A.4 Turn the ventilator ON.
6.25A.5 Adjust the Inspiratory Flow control on the ventilator to Medium.
6.25A.6 Adjust the fine flow control of the Ped Bellows attachment fully
clockwise.
6.25A.7 Set the O
2
to 3 l/min.
6.25A.8 Occlude the 15mm patient side of the Y-piece and fill the bellows by
pressing the O
2
Flush button.
6.25A.9 Observe the absorber breathing pressure gauge as the ventilator cycles.
6.25A.10 What is the pressure when the bellows completes its downward travel?
___cmH
2
O >30
6.25A.11 What is the pressure when the bellows completes its upward travel?
___cmH
2
O 3
6.25A.12 Verify that with the Pediatric Bellows Fine Flow Control turned fully
counter-clockwise the bellows does not collapse during inspiration.
6.25A.13 Installa3Lbreathing bag onto the patient circuit Y-piece.
6.25A.14 Fill the bellows by pressing the O
2
Flush button.
6.25A.15 Adjust the O
2
flow to 10 l/min.
( ) 6.25A.16 What is the PEEP? ___cmH
2
O(3)
6.25A.17 Adjust the O
2
flow to 500 ml/min.
( ) 6.25A.18 Does the ventilator deliver the full tidal volume during 5 consecutive
cycles? ___(Y)
6.25A.19 Set the ventilator frequency to 20 BPM.
Rev. Y