ADJUSTMENT AND CALIBRATION PROCEDURES (continued) NM2B
CONNECTION
ON SENSOR
INTERFACE
PANEL
ABSORBER
ASSEMBLY
TOP VIEW
TEST
FIXTURE
BREATHING
PRESSURE
SV20322
Figure 5-7: BAROMED CALIBRATION SETUP AND SCREEN
CALIBRATION MENU O2MED
OFFSET O2 CELL A 0138 0127
OFFSET O2 CELL B 0138 0129
BAROMED
OXYGEN OFFSET READINGS STORED
PRESSURE SPAN (60 cm H2O) 0243 0481
EXIT
TRACE
TREND
Rev. A
5-22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS