NM2B PMC PROCEDURE (continued)
6.16.17 What is the high oxygen alarm default? ___ % (100)
6.16.18 Verify that the high alarm limit has a range from 100 to 22%.
6.16.19 Set the high alarm to 95.
6.16.20 the alarm CRT shall display "% OXYGEN HI" under the ADVISORY
column.
6.16.21 Return the high alarm to 100.
6.16.22 The "% OXYGEN HI" message will now disappear.
( ) 6.16.23 Within three minutes, what is the oxygen concentration? ___% (97-100)
Rev. Y
6-39