Technical data
Dräger Polytron 8200/8310 29
12.1 Torques
(The torques are valid for the aluminum and stainless steel
316 variants)
If an optional e-Box is being used, fasten the device securely
to the e-Box with 4 screws and a tightening torque of 8 Nm.
12.2 Tightening torque and cable size for field
wiring terminals
12.3 Factory settings
12.3.1 Alterable settings for the Polytron 8200/8310
Environmental conditions Specifications for the sensor: see sensor data sheet
Pressure 700 to 1300 hPa (20.7 to 38.4 in. Hg)
Humidity 0 to 100 % R. H., non-condensing
Temperature
DrägerSensor DD, LC no relay
DrägerSensor DD, LC with relay
DrägerSensor IR no relay
DrägerSensor IR with relay
-40 to +80 °C (-40 to +176 °F)
-40 to +70 °C (-40 to +158 °F)
-40 to +65 °C (-40 to +149 °F)
-40 to +65 °C (-40 to +149 °F)
Environmental influences See respective sensor data sheets.
Part Torque Lb. In. Torque Nm
Housing cover min. 266 min. 30
Sensors min. 266 min. 30
Plug min. 266 min. 30
Cable bushing min. 443 min. 50
Device to e-Box 71 8
Electronic
Torque Lb. In. Cable size in AWG Cable size in mm
2
All field wiring
terminals
4.4–7.0
(0.5–0.8 Nm)
20–12 0.5–2.5
Standard setting
Menu DrägerSensor
DD, IR
DrägerSensor
LC
A1 Alarm 20 %LEL 2.0 %LEL
A2 Alarm 40 %LEL 4.0 %LEL
A1 Relay alarm:
Direction
Rising Rising
A2 Relay Alarm:
Direction
Rising Rising
A1 latching mode Non-latching Non-latching
A2 latching mode Latching Latching
A1 Acknowledgement
mode
Acknowledgeable Acknowledgeable
A2 Acknowledgement
mode
Not
acknowledgeable
Not
acknowledgeable
A1 enrg / wired in e-Box Energized / NO Energized / NO
A2 enrg / wired in e-Box Energized / NO Energized / NO
Maintenance password 0000 0000
LCD setting on on
Configuration/SIL
password
_ _ _2 _ _ _2
SIL status Off Off
Language English English
Fault current 1.2 mA 1.2 mA
Warning current 3.0 mA 3.0 mA
Maintenance current 3.4 mA 3.4 mA
HART address 0 0
Auto-cal. Off Off
Calibration interval 360 360
Function key Faults Faults