Testing
Symbol "Battery warning 2"
In this case the red LEDs flash constantly. On power up
5 short beeps will sound and the display will show the
"Battery warning 2" icon.
Sentinel automatically powers off and any further operation
is impossible.
Releasing Air from the Breathing Circuit
Remove sealing cap R 33 588 from the plug-in coupling
and wait until Sentinel shows the zero pressure.
1 Replace the sealing cap R 33 588 on the plug-in coupling.
Switching off Sentinel
Simultaneously press the right and left hand button until a
single sharp audible "Bleep" sounds.
Release buttons.
For 3 seconds Sentinel shows the battery status.
Sentinel switches off.
WARNING
If battery warning 2 is displayed, the battery must be repla-
ced before any new operation.
Failure to follow this warning could lead to death or serious
injury.
02621507.eps
bar
5x Bleep
13121507.eps
1
11321507.eps
bar
1x Bleep
bar
04621507.eps
Page 48