Storage
Completion of Testing
Closing the Protective Cover
(see “Closing the Protective Cover” on page 17).
Storage
After fitting the CO
2
absorber in the operational PSS BG 4,
the shelf life specified in the Maintenance Intervals (see
page 65) should be observed. Do not exceed the 'use by' date.
Breathing bag, breathing hose and relief valve must be
installed ready for operation and the plug-in coupling must be
fitted with a sealing cap R 33 588.
Store the PSS BG 4 ready for use in a dark, cool, dry, dirt-
free, and dust-free place.
If PSS BG 4 is stored upright, be sure it will not fall over.
Avoid exposure to UV rays and ozone.
Observe the national standards for elastomers.
DANGER
The PSS BG 4 is only ready for use when all the above pre-
parations are complete!
If one test failed, the PSS BG 4 must not be used. In this
case , it must be serviced by trained personnel or by Dräger
Safety prior to the next use.
Failure to follow this warning could lead to death or serious
injury.