Camera (cont’d.)
Fault Cause Remedy
Battery does not charge in the truck charger
while attached to the camera.
Safety circuit of the battery may be tripped. Battery needs to be re-charged while
disconnected from the camera. Take battery
out of the camera and charge with a desktop
charger or use battery-only charging slot in
the truck charger. If charging does not work,
contact Dräger Safety.
Battery is put in the battery slot of the
desktop charger or truck-mounted charger
and the red LED on the charger blinks.
Battery is out of the valid charging
temperature range (32
o
F to 95
o
F; 0
o
C to
35
o
C).
Ensure that battery and charger are within the
valid charging temperature range, then
charging should work normally.
Battery is over-discharged. Leave battery on the charger until charging is
completed and the green LED comes on.