73
Incorrect sensor test The sensor self-test has detected a fault in a
sensor.
Switch the instrument off and then on again, or
remove and reconnect the power pack. If the fault
is still displayed, contact DrägerService.
Low IR signal The instrument has detected a fault in the
IR sensor.
#126: IR signal too low.
Switch the instrument off and then on again, or
remove and reconnect the power pack.
Change IR sensor, page 86
If the fault is still displayed, contact
DrägerService.
High IR signal The instrument has detected a fault in the
IR sensor.
#127: IR signal too high.
Switch the instrument off and then on again, or
remove and reconnect the power pack.
Change IR sensor, page 86
If the fault is still displayed, contact
DrägerService.
Sensitivity low. Please check lamp #132: Span falls short of warning setpoint Clean the lamp of the PID sensor, see sensor
data sheet. If the device continues to display this
error, replace the lamp of the PID sensor, see
sensor data sheet.
Error during sensor test # 108: Bump test failed Repeat bump test, ensure that gas is continu-
ously supplied during the test. If required, read-
just the sensors.
Rectify the channel or instrument fault and repeat
the Bumptest.
Automatic Bumptest does not start Dräger X-am 7000 were not configured
according to Dräger CC Vision
Configure the Dräger X-am 7000.
There is no activated printer with correctly
positioned IR interface in front of the
BumpTest Station
Switch on the printer and position the
IR interface of the printer in front of the fiber
optic cable of the BumpTest Station.
The manual Bumptest cannot be started via the
"Quick Menu"
One of the channels activated for the Bumptest is
in WarmUp or Error mode or an instrument fault
has occurred
Wait until the WarmUp time has elapsed, rectify
the fault and restart the Bumptest.
Fault Cause Remedy