10 Handheld Pulse Oximeter (Reorder No. 36403)
1. INTRODUCTION (Continued)
Any condition that restricts blood ow, such as use of
a blood pressure cuff or extremes in systemic vascular
resistance, may cause a failure to determine accurate
pulse rate and SpO
2
readings.
Remove ngernail polish or articial ngernails before
applying SpO
2
sensors. Fingernail polish or articial
ngernails may cause inaccurate SpO
2
readings.
Optical cross-talk can occur when two or more sensors are
located in adjoining areas. It can be eliminated by covering
each site with opaque material. Optical cross-talk may
adversely affect the accuracy of the SpO
2
readings.
Obstructions or dirt on the sensor’s red light or detector
may cause a sensor failure. Make sure there are no
obstructions and the sensor is clean.
For routine equipment maintenance, please refer to the
service procedures at the associated section as indicated
in the manual.
As to the other concerns for attention, please carefully look
through the specic chapter in enclosed instructions.
This manual describes all features and options. This model
may not have all accessories mentioned in this manual.
Federal Law restricts this device to sale by or on the order
of a physician.