2. Attach the Encoder Housing to the long plate (plate #1) using two 1.25" long
SHCS, two 5/8" long SHCS, 4 washers & 4 lock washers (washers first then
lock washers, i.e. lock washers under screw head).
Note for Pneumatic Lift Users: With the encoder arm in the down position,
lower the pneumatic lift. Move the arm up and down to check for clearance
with the pneumatic lift bucket. Clearance may be increased a small amount by
rotating the encoder housing before tightening the screws. If no clearance exists
and the arm hits the pneumatic lift bucket, contact Dynisco Polymer Test
Systems before proceeding.
Tighten the screws
3. Plug the phone jack cable into the encoder housing.
4. Move the encoder arm into the down position
5. Turn power on
6. Enable the Encoder by pressing the SETUP key. Browse the options until
you reach B FLAG and press EDIT. Press Quick Key so that ENCODR is
over the SELECT Quick Key. SELECT encoder and press ENTER to accept.
Press ESC. To back out to main menu.
7. Press the SYS key and select TEST. Browse the menu and select Digital
Encoder Test. The display shows current position and should change when the
arm is moved. Press QUIT to end procedure. ESC out to main menu.
8. See section entitled for detailed instructions on the variety of uses and
programming available with the encoder option.
Programming for use with the Digital
Encoder
Method B: Setup
Enable the Digital Encoder
Auto Flag ON/OFF
This procedure assumes you are already set up to perform a Method B test as
explained in the section entitled Method B Run”. Make sure Method B is
selected in the EDIT mode.
Press the SETUP key and browse the options until you reach B FLAG and
press EDIT. Press Quick Key so that ENCODR is over the SELECT Quick
Key. SELECT encoder and press ENTER to accept. Press ESC. twice to back
out to main menu. Enter the EDIT mode.
Press Quick Key until AUTO FLAG appears. Select ON to enable the
instrument to automatically determine one flag length based on real time flow
rates during the actual run. ASTM standard flag lengths will be used. No data
entry will be required. Auto flag is the easiest mode, it handles all the flag
details, unless you have a detailed test method use this option.
Select NO then you must enter more information in the options below about the
number of flags and their length. Press ENTER for next prompt.
Series 4000 Melt Indexer Operation Manual The Digital Encoder: Use and Calibration