EasyManuals Logo

E Instruments E4500 User Manual

E Instruments E4500
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
96
K000000000EJ 023826A0 100714
07/08/14
10:00
New piping
Stabilization
o
oo
o
hPa
100.00
P diff
hPa
100.00
∆P 1 min
00:15:00
Wait time
Automatically
Automatically,
after 15 minutes.
07/08/14
10:15
New piping
Test
o
oo
o
hPa
100.00
P1
hPa
100.00
P2
hPa
0.00
∆P
00:15:00
Wait time
o
oo
o
07/08/14
10:15
New piping
Test
F1 F2
hPa
100.00
P1
hPa
100.00
P2
hPa
0.00
∆P
00:15:00
Wait time
WARNING
Tightness test ongoing.
Stop?
F1: stop
F2: cancel
F1
07/08/14
10:00
New piping
Pressurization
P diff
hPa
0.06
O
NOTE: If, while configuring the tightness test the automatic printing mode has been selected, the tightness test is
printed automatically.
Instead, if the manual printing mode has been selected (exemplified case), at the end of the tightness test the
results are displayed and they can be saved and/or printed.
In this case proceed as follows:
Pressure of the system.
Pressure variation in the last minute.
Waiting time to end the stabilization phase.
The stabilization phase can be interrupted in any moment.

Table of Contents

Other manuals for E Instruments E4500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the E Instruments E4500 and is the answer not in the manual?

E Instruments E4500 Specifications

General IconGeneral
BrandE Instruments
ModelE4500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals