User manual SEFELEC 5x Series 156 v1.06
CTRLOUT (Output C7-5)
Validation contact for the taking of control of the device by the PLC.
EOT (Output C7-7)
End of Test contact
PLCPASS (Output C7-8)
Test result contact: PASS
PLCFAIL (Output C7-6)
Test result contact: FAIL
ERROR (Output C7-9)
Indicates whether an error occurred during the test; outputs SR0 to SR3 give the type of
error
SR0 to SR3 (Inputs C8-5,6,7,8)
Coding of the type of error according to the following table:
SR3
SR2
SR1
SR0
Type of error
0
0
0
0
Open Safety loop
0
0
0
1
Saturation
0
0
1
0
Over range
0
0
1
1
Charge
0
1
0
0
Voltage error
-
-
-
-
1
1
1
1
….
IN_OUT_0-10V_0 (Output C6-1)
Analogue output 0-10VDC, configurable
IN_OUT_0-10V_1 (Output C6-2)
Analogue output 0-10VDC, configurable
IN_0-10V_2 (Input C8-9)
Analogue input 0-10VDC, configurable
IN_ 0-10V_3 (Input C7-3)
Analogue input 0-10VDC, configurable