XS series operating manual 
 
XS user manual v0.55  - 122 - 
Common queries 
 
*STB? : 
return in hexadecimal format a <NR1> which is the value of "STB". (i.e. : "#H80") 
 
 
0 = Interlock opened 
1 = Interlock closed 
0 = No error 
1 = Error (voltage not correct,.....) 
0 = End of test 
1 = Test running 
0 = Test fail 
1 = Test pass 
Logical OR of the ESR register bits 
Logical OR of the STB register bits 
 
*SRE? : 
return  in  hexadecimal  format  a  <NR1>  which  is  the  value  of  "SRE".  (STB  register 
mask) 
 
*ESR? : 
return in hexadecimal format a <NR1> which is the value of "ESR". 
 
1 = Dialog error type 2 (Numerical value out of range, out of context command) 
1 = Dialog error type 1 (incorrect command) 
 
NOTE : the bits are reset after the reading of the byte by the *ESR? command  
 
*ESE? : 
return in hexadecimal format a <NR1> which is the value of "ESE". (ESR register mask) 
 
*LRN? : 
This device query allows the programmer to receive a response message unit that informs 
the  programmer  on  the  current  device  state  (function's  running  and  current  test 
parameters)  and  may  be  later  used  as  program  message  unit  elements  to  place  the 
device in the state it was. 
 
*IDN? :  
Allows the identification of the unit. The message is as follows: 
    <field1>,<field2>,<field3>,<field4> 
    field1 : Manufacturer name  = Sefelec 
    field2 : Unit reference = CMG30, DMG50, DMG500, RMG50, MMG500, SMG50... 
    field3 : Serial number = 0 (not used) 
    field4 : Software revision = VERSION 1.60 
   
NOTE : the *IDN? command must be used before selecting a function (from the start display)