9.1 Service
Pump exchange:
https://youtu.be/lKHx5F-hnwY
CAUTION!
Dispenser care
Inadequate dispenser hygiene can result in microbial colonisation of the
dispenser system. The most frequent germs are aerobic sporulating
organisms, Corynebacteria, various cocci and pseudomonas. Hand
sanitizer and soap residues in the hose or on the nozzle can also cause
problems.
NOTICE!
For hygienic reasons, the dispenser pumps are designed as disposable
pumps which must be replaced at regular intervals every 3 - 4 months, or
after 4 canister changes at the latest.
The pumps must not be reconditioned, but must be replaced!
Label pump with date prior to using it for first time!
When used for the first time or after a long period of use, the tube may be
stuck in the pump head at the point where the rollers pinch it off. This can
be remedied by moving the visible part of the gear on the back of the pump
head back and forth several times.
Removal of the pump
Fig. 6: Remove the pump
1.
Open dispenser with key (see
Ä
Chapter 3 ‘Scope of delivery’ on page 17).
2. Take the nozzle out of the bracket (Pos. 1).
3. Turn the pump (pos. 3) clockwise (approx. 20°) until the lock is released.
4. Remove the pump.
Service
32MAN048961, Rev. 2-09.2020