12
8.0 SERVICE
8.1 Regular Inspection
1. Inspect the pump squeeze tube. If visible worn or
cracking, replace the squeeze tube even if it is working
properly at the time.
NOTE: Properly lubricate the tubing with the silicon
lubricant provided.
2. Inspect all tubing connections for leaks, cracks or loose

8.2 Squeeze Tube Replacement Instructions
1. Loosen tube nuts and pick-up discharge ends of squeeze
tube.
2. Remove the plastic cover from the face of the peristaltic
pump.
3. Pull both ends of the squeeze tube away from pump
body until tubes slips over the rollers.
4. Clean inside of pump body, wiping to remove particle
matter and accumulation of old lubricant.
5. Inspect the roller assembly closely for signs of wear or
cracking.
 

driver and rotate the roller assembly clock-wise to assist
in the placement of the tubing.
 
silicon grease lubricant.
 
NOT over tighten since plastic can stress crack if excess
of force is applied.
 
to check proper function of the new squeeze tube.