Countermeasures For Bad Lead Connection
After turning on the power switch, enter the ECG test to check waveforms of all leads.
On this menu, operators can check waveforms of all leads and their noise, and may
see “Lead Fault” message displayed on the LCD screen.
NOTE
- Message will be displayed when L/F is set to ON in System Setup, and will
NOT be displayed when set to OFF.
-
“Lead Fault” message will appear, and a buzzer alarm will go off when lead
fault occurs during printing.
The following two cases could cause a lead fault error.
- First case, leads are detached from the patient’s body. In this case, re-attach leads
normally.
- Second case, signal conductivity between leads and the patient’s body is low. In
this case, replace the electrodes.
If the two above cases have been performed and the operator is still experiencing
noise or lead fault conditions, the patient cable may be faulty. Please contact the
Medical Econet GmbH service center.
Lead Fault! All Leads.