WARNING
When a defibrillator is used, do not touch the patient cable or apparatus.
WARNING
When you are connecting to the electrodes or the patient cable, make sure
connectors never come into contact with any other conductive parts, or are
grounded. In particular, make sure that all of the ECG electrodes are attached
to the patient, to prevent them from contacting conductive parts or contact
with ground.
WARNING
Use the provided ECG patient cable for measuring ECG in the operating room.
The patient cable cannot be used for measuring respiration.
WARNING
Do not use the Cardio M Plus in combination with any Electro-Surgical (ES)
equipment.
WARNING
Users must use the same type of electrode or any other electrodes with
biocompatibility certification proven by one of the international standards.
WARNING
Cardio M Plus must be operated under the supervision of a health care
professional, especially when it is used on a pacemaker-treated patient.